General Specifications
DIMENSIONS AND WEIGHT
• L x W x H: 240 x 190 x 30 mm
• Net Weight: 600 gr. without battery
• Body material: Plastic cabinet
TEST TYPES
Pure Tone Audiometry, Automatic Audiometry
Patient controlled, according to ISO 8253-1 and
standard Hughson Westlake procedure (selectable
on control panel)
Audiometer Class 4 according to IEC 60645-1
TALK FORWARD
• Built-in talk forward microphone 50 – 110 dB SPL
(adjustable on control panel)
STORE FUNCTION
• Dedicated Store button that enables to store
measurements into the internal memory
(up to 50 patients).
The patient will be identied by an ID. number.
• Stored data can be viewed on the built-in display
or transferred to the PC using MDS software.
PC INTERFACE
• Resonance MDS Management Data Suite, with
reporting and printing features.
Audiometry Specification
RANGE
• Frequency range (AC):
125 – 8000 Hz (DD45)
125 – 8000 Hz (Insert Phones)
INPUTS
• Tone
• Warble Tone +/- 5%, 5 Hz
(true sine wave frequency modulation)
OUTPUTS
• Left and Right
TONE PRESENTATION
• Manual or Reverse (selected on Setup Menu)
• Pulsed at 0.5,1 or 2 Hz – (selected on Setup Menu)
or Continuos
FREQUENCIES AND INTENSITIES WITH DD45
All the frequencies can be enabled or disabled by the
Setup menu except 1000 Hz.
Frequency (Hz) | Intensity (dB HL) |
---|---|
125 | 65 |
250 | 85 |
500 | 100 |
750 | 100 |
1000 | 100 |
1500 | 100 |
2000 | 100 |
3000 | 100 |
4000 | 100 |
6000 | 100 |
8000 | 85 |
LEVEL STEPS
1, 2 or 5 dB (selected on Setup Menu)
DISTORTION
• 0.3% typical
• below 1% at full intensity
RISE/FALL TIMES
40 msec.